Medtronic, a developer, manufacturer, and seller of medical therapy devices, has initiated a voluntary recall of its MiniMed insulin pumps, specifically the 600 and 700 series. The recall was prompted by concerns over a potentially shortened battery life, which could affect the device’s function and the user’s health. The action was publicly announced on July 31 as a field action to global customers, alerting them to carefully monitor their MiniMed pump’s built-in alarms and alerts, which help track battery status.
The decision to analyze and eventually recall certain insulin pumps arose after Medtronic received customer feedback reporting that physical impacts such as bumps or drops could lead to a reduced battery lifespan. These impacts potentially damage the pump’s internal electrical components, which might not only diminish the battery life but also alter the accuracy of battery life notifications compared to what is stated in the User Guide. This issue could lead the pump to stop delivering insulin much earlier than expected, posing severe health risks such as hyperglycemia or diabetic ketoacidosis (DKA). Both conditions are serious and might require urgent medical intervention.
Medtronic’s statement highlighted the gravity of the situation, reporting that there were 170 incidences of hyperglycemia with blood sugar levels exceeding 400 mg/dL and 11 instances of diabetic ketoacidosis possibly linked to the battery life problem in the U.S., recorded between January 2023 and September 2024. To address these risks, the U.S. Food and Drug Administration (FDA) has classified this issue as a Class I recall. This classification is the most severe category, indicating that continued use of the affected pumps poses a reasonable probability of causing significant injury or death.
Despite the battery life issue, Medtronic reassured users that the insulin pumps would still trigger battery alerts and alarms. However, these notifications may not display at the usual times, potentially reducing the warning time before the battery fails and the pump ceases insulin delivery. To mitigate any potential disruptions in insulin therapy, Medtronic has advised users to replace the battery immediately when the “Low Battery Pump” alarm activates. Additionally, users are recommended to carry an extra set of batteries and have an alternative diabetes management plan ready as a precautionary measure.
Medtronic urged all users, irrespective of whether their device was dropped or impacted physically, to contact the company for assistance if they notice a significant reduction in the device’s battery life. This proactive approach aims to determine if a replacement pump is necessary to ensure continuous and safe insulin delivery.
The recall has stirred concerns among users of the MiniMed 600 and 700 series insulin pumps, underscoring the importance of robust device manufacturing and quality assurance processes in the production of medical devices. Medtronic, based in Dublin, is managing the recall and customer communications, ensuring that those affected by the battery issue are informed and equipped to handle the potential challenges posed by this problem. The initiative reflects Medtronic’s commitment to patient safety and responsiveness to product performance issues.